FAQ
Frequently asked questions on W&H products and helpful descriptions on the topic of hygiene and care.
Koblinger
-
Roto Quick (5)
-
Can I use all instruments which fitted the old Roto Quick coupling (924, ...) with the new RQ-24 too?
Yes. All existing and new instruments can be used with couplings from both the old and the new generations.
-
How does the new W&H Click & Pull system work? How do I remove a turbine?
Use your thumb and index finger to retract the coupling's retention sleeve. The turbine can now be removed from the coupling without any problems. Pulling forcefully on the dental turbine can cause damage. If excessive force is required to remove the turbine, verify that the retention sleeve on the coupling can be moved without any problems.
-
Why is a 6-hole connection required for the use of an LED turbine?
A 6-hole connection for the turbine tubing is necessary to supply the turbine with electricity via electrical contacts. Other connections do not provide a power supply.
-
Can the W&H coupling be thermodisinfected and sterilized?
No, thermodisinfection of the coupling is not permitted; sterilization in steam sterilizers can be applied at a temperature of 135°C. Hot air sterilization and/or cold sterilization (placing in liquids) are not approved.
-
Are there other couplings than the Roto Quick available?
Yes. W&H also offers the "RM-34 LED" for Multiflex® connection.
-
Can I use all instruments which fitted the old Roto Quick coupling (924, ...) with the new RQ-24 too?
Tilbehør
Kirurgiske save
-
How to dismantle the handpiece
See detailed description in user manual or Video.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
Osstell stabilitetskontrol
-
Osstell Beacon (2)
-
Where can I buy the SmartPegs?
Please contact your local W&H partner
The sale of SmartPegs will be defined per country
The surgeon will receive 5 free SmartPegs. A voucher will be provided with the delivery of the product. -
What does ISQ stand for?
The abbreviation ISQ stands for Implant Stability Quotient. It is a value between 1 and 100 and gives the surgeon an insight of the implant stability after placing the implant. With the ISQ value the surgeon can monitor the osseointegration.
-
Where can I buy the SmartPegs?
Luftscaler
-
Proxeo (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
Elektrisk motor
-
Elektrisk Motor (EM-12 L) (10)
-
How do I connect the electric motor to my unit as an Add-on?
To connect the electric motor to your dental unit as an Add-on you require a free turbine hose on your dental unit and the possibility of connecting the control to the mains via the included power supply unit. You can control the electric motor using the foot control of your dental unit.
-
Can I sterilize my electric motor in a sterilizer?
Yes, if your electric motor features the symbol “Sterilizable up to the stated temperature”.
-
Can my electric motor be reprocessed in a thermo washer disinfector?
No, that is not possible.
-
Does the Built-in version of the electric motor require maintenance?
The Built-in electronic component does not require separate maintenance. However, regular servicing together with the rest of the dental unit should be performed at the intervals prescribed by the pertinent legislation. This makes it possible to guarantee proper function and safety. More detailed information on this topic can be found in the “Service” section of the respective instructions for use.
-
Does the Add-on version of the electric motor require maintenance?
The electric motor as an Add-on version including accessories requires a regular check once every three years unless shorter intervals are prescribed by law. The regular service must only be performed by an authorised W&H service partner. More detailed information on this topic can be found in the “Service” section of the instructions for use.
-
Does the electric motor need to be lubricated with oil?
No, the motor bearings require no maintenance. Any additional oil service would reduce the service life of the bearings.
-
Should I always call my W&H service partner in advance before sending devices/instruments in for repair?
If you are not completely sure of the cause of the problem, it is better to contact your W&H service partner first and speak to a technician. It is sometimes possible to resolve malfunctions together with the technician and decide what needs to be sent in for repair.
-
What is a brushless electric motor and what benefits does it offer me as a user?
The brushless electric motor is considerably less susceptible to wear than a brush motor, which in turn translates to minimised service and maintenance work. This also means less downtime for the user when servicing proves necessary.
-
Which contra-angle handpiece series is recommended for the electric motor?
The W&H Synea series – the glass rod elements are optimally suited for use together with the motor and its integrated LED+.
-
How can I integrate the electric motor into my unit as a Built-in version?
The electric motor can be integrated into a dental unit by selected dental unit manufacturers or their service technicians.
-
How do I connect the electric motor to my unit as an Add-on?
-
Elektrisk Motor (EM-E6) (2)
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
No. You can purchase an Allegra LED G contra-angle which has its own light generator and a LED. No further investment necessary.
-
Does the electric motor need to be greased?
No, the motor bearings require no maintenance. Any additional greasing would reduce the service life of the bearings.
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
-
LED opgraderingssæt (2)
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
No. You can purchase an Allegra LED G contra-angle which has its own light generator and a LED. No further investment necessary.
-
Does the electric motor need to be greased?
No, the motor bearings require no maintenance. Any additional greasing would reduce the service life of the bearings.
-
I've got a micromotor without light. Do I have to change the micromotor and the contra-angle to have light?
Tilbehør
-
Service Oil F1 (6)
-
Why is W&H oil better than other lubricants?
Extensive research and testing produced a formula which best covers the extreme demands of dental instruments. It consists of an all-synthetic special oil of the highest purity supplemented by the specially developed additive package.
-
Can W&H Service Oil F1 be sterilized?
Yes, W&H has checked the sterilizability precisely. W&H Service Oil F1 was not found to influence the sterilization of the instruments in any way.
-
Up to what maximum temperature is the oil stable without problems?
W&H Service Oil F1 has longterm stability between –30 oC and +160 oC.
-
Which aerosol is used in the W&H F1 spray can and what properties does it have?
A mixture of propane and butane (standard aerosol in all spray cans). It contains no CFCs and as such is environmentally-friendly.
-
How should I dispose of empty cans?
In accordance with the national regulations for aerosol cans. More information on disposal can be found on the safety datasheet.
-
Can I use any oil to lubricate my W&H instruments?
W&H recommends that W&H instruments are lubricated with W&H oil which has been designed and developed to enhance the performance of your W&H instrument.
-
Why is W&H oil better than other lubricants?
-
Seal2 (1)
-
Should I pouch/wrap my instruments and how long can I store them after sterilization?
To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer.
-
Should I pouch/wrap my instruments and how long can I store them after sterilization?
Hånd- og vinkelstykker
-
Synea (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
Hånd- og vinkelstykker
-
Endea (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
-
Endo NiTi (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
-
Endea Endo Cursor (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
Kirurgi maskiner
-
Implantmed (34)
-
How can I adjust the torque?
The torque can be adjusted in P4 & P5 only. In- or decreaseing is done via either "+" or "-" button. A detailed description (factory settings, adjstable range,…) can be found in the user manual.
-
Can handpieces from another manufacturer be used on a W&H surgical motor (implantmed / elcomed)?
For applications which do not require accurate torque (drilling, grinding,…), any handpiece with ISO 3964 connection can be used. Please be informed that hand pieces from other manufacturers might not be able to handle the torque of your unit/motor and therefore might get damaged.
-
Can other brands of tubing be used?
Not recommended. Due to different specifications (material, dimensions) it could cause problems like leakage, insufficient cooling or even damage the unit. For propper tubing see "correlation of tubing" or user manual.
-
How to insert irrigation tube?
Please find a detailed description in the user manual at chapter "Starting operation - General".
-
How can the unit / motor be cleaned / sterilized?
Please find a detailed description in the user manual at chapter "Hygiene and Maintenance".
-
Is it possible to set the motor speed with the program for positioning an implant?
No, the speed in program P4 is 15 rpm when in forward operation and 30 rpm in reverse operation. In program P5 (thread cutter function) the speed is 20 rpm in both forward and reverse operation and only the torque can be set.
-
Why doesn't the light on my LED G contra-angle handpiece work when I am positioning an implant?
The generator in the contra-angle handpiece requires at least 300 rpm to produce enough electricity for the LED. The rotary speed is between 15 and 50 rpm when positioning an implant.
-
Can my Implantmed/Elcomed motor be processed via thermal washer disinfector?
Yes, if the thermal washer disinfector has a drying cycle and the motor bears the sign that it is "Thermo washer disinfectable".
-
Can my Implantmed/Elcomed motor be processed in a vacuum sterilizer with the protection cap in place.
No. The new generation of motors have to be sterilized without cap.
-
Does my W&H unit need to be serviced.
Regular servicing of function and safety including the accessories is necessary and should be carried out at least once every three years, unless shorter intervals are prescribed by law. Detailed information can be found in the user manual at chapter "Servicing".
-
What types of tubing can I use on my device?
For propper tubing see "correlation of tubing" or user manual.
-
Does the motor need to be greased?
No, the Implantmed/Elcomed motor does not need greasing. The motor bearings require no maintenance. Any additional greasing would reduce the service life of the bearings.
-
Can previous Implantmed units be retrofitted with the wireless foot control
It depends but basically yes. The following units can be retrofitted
Implantmed SI-923 REF16929000
Implantmed SI-915 REF16929001
Elcomed SA-310 REF15933100/15933101/15933102/15933103
Please contact your local W&H Partner to find out if the wireless foot control is already registered in your country -
How does it work to wirelessly control the Implantmed and the Piezomed?
W&H introduces a new concept in oral surgery and implantology. Via dongles (receiver), which can be connected to the unit by the doctor, the wireless foot control S-NW communicates with the surgical units of W&H.
-
Are old cable foot controls compatible with the new unit or vice versa?
No. The new cable foot control S-N2 is based on the industrial CAN standard. So connecting the old foot control S-N1 is not possible.
-
Can our customers disassemble the new contra-angles
Yes. At W&H the letter "S" always communicates, that the surgical contra-angle is disassemble. The letter "I" communicates, that the surgical contra-angle cannot be disassembled (e.g. WS-75 L vs. WI-75)
-
Can the new contra-angles be attached to old motors?
No. In order to deliver the best balance we shortened the inner coupling length. Therefore contra-angles with only an L suffix cannot be attached to old motors
-
Can current contra-angles be attached to new motors?
Yes. All W&H surgical contra-angles and handpieces can be attached to the new motor EM-19.
-
What does the LC in EM-19 LC stand for, is there a LED in the motor?
The LC stands for Light Contacts. This means that we have integrated an electrical connection into our new motors. The Light Contacts deliver power to our new contra-angles to offer our customers the best light at any rotation speed. For oral surgery handpieces from W&H, the LED remains in the head of the contra-angles
-
Are previous Implantmed motors compatible with the new Implantmed or vice versa?
No. The new motor has a new technological basis and is the shortest available
-
Can the surgeon upgrade the unit by themself or does the unit need technical service?
The surgeon or dental nurse can easily attach the module to the right side of the Implantmed with the help of the provided quick manual. The module is plug-n-play via a USB connection.
-
Is every new Implantmed compatible with the W&H Osstell ISQ module?
Yes. All control units SI-1023, SI-1015 and SI-1010 can be upgraded with the W&H Osstell ISQ module
-
What’s the difference between torque and ISQ?
Torque is a one time, static measurement at the time of placement of the implant and cannot be repeated later on in a non-invasive way. Osstell ISQ measures the lateral stability of the implant and the measurements can be repeated in a non-invasive, dynamic way to monitor the development of osseointegration.
-
Where can I buy the SmartPegs?
Please contact your local W&H partner
The sale of SmartPegs will be defined per country
The surgeon will receive 5 free SmartPegs. A voucher will be provided with the delivery of the product. -
How does the measurement of the ISQ work?
> To measure the ISQ of an implant, you first screw a so called SmartPeg into the placed implant. The SmartPeg, with its magnet on top, works like a small tuning fork. The magnet on the SmartPeg is ‘hit’ with magnetic pulses, from the probe, which makes the SmartPeg vibrate. Due to the stiffness in the interface between the implant surface and the bone the SmartPeg will vibrate accordingly. The more dense the bone is the higher stability and the higher ISQ value.
-
What does ISQ stand for?
The abbreviation ISQ stands for Implant Stability Quotient. It is a value between 1 and 100 and gives the surgeon an insight of the implant stability after placing the implant. With the ISQ value the surgeon can monitor the osseointegration.
-
Piezomed module: Does the module also work with OEM devices?
Yes, but different software updates are necessary.
-
Piezomed module: Can devices be retrofitted on the market?
Yes, all SI-10XX devices.
-
Piezomed module: Are the handpieces of the modules compatible with the SA-320?
No, not compatible.
-
Piezomed module: What are the main differences between the Piezomed Plus module
and Piezomed Classic module?
Features Piezomed Plus module Piezomed Classic module Power 24 watts 18 watts Light LED no light Tip detection yes no Colour black white Cable length 1.8 m and 3.5 m 1.8 m -
Piezomed module: Compatibility of the handpieces with Piezomed Plus module
and Piezomed Classic module.
Handpieces Piezomed Plus module Piezomed Classic module SA-40 yes yes SA-40 L (light version) yes no SA-320 no no -
Piezomed module: Is it possible to upgrade every existing product on the market?
Yes, Implantmed Plus from serial no. SN01001.
-
Can I use tips from competitors?
No, thanks to the automatic tip detection and the special design, W&H offers a closed system.
-
Piezomed module: Can I use the same Piezomed tips for SA-320 and the new modules as well?
Yes.
-
How can I adjust the torque?
-
Piezomed (15)
-
How long is the useful service life of the LED?
The user must replace the LED socket as soon as the protective coating on the LED discolours and the luminous power is impaired as a result. The length of time this takes can differ depending on the type of steriliser or sterilization process.
-
Can Piezomed instruments also be used on third-party devices?
No
a. Different thread
b. Third-party device cannot find the correct resonance point – instrument fails to oscillate -
How long do the instruments last and when should they be replaced?
The useful service life of the instruments is determined by the length of time they are used reprocessing and sterilization cycles have only marginal influence.
a. Instruments need to be replaced
- If there is a loss of power while working
- Where there is visible damage on the working part or the shank of the instrument.
b. Saws must be replaced if
- Teeth are broken off
- Teeth are worn
c. Diamond-coated instruments must be replaced
- As soon as the diamond coating is worn
-
How are the instruments prepared for cleaning?
a. Preparation in the ultrasonic bath:
- Place the instruments in the instrument tray and lower into the ultrasonic bath.
- Use cleaning agents and disinfectants suitable for hand instruments (probes, mirrors, forceps, etc.).
DO NOT USE a bur bath, as this is very aggressive.
- After cleaning in the ultrasonic bath, rinse adequately with water to remove any remnants of the cleaning agent and disinfectant from the coolant channels.
- Blow dry with compressed air after rinsing
- Replace dried instruments in the instrument tray and package for sterilization
b. Mechanical preparation in the cleaning unit and disinfector
- Use the cleaning adapter for cleaning units and disinfectors (see Instructions for Use – Accessories)
- After preparation in the cleaning unit and disinfector, check that the coolant channels are dry and, if necessary, dry again with compressed air. -
Can third-party instruments be used on the Piezomed?
No
a. Different thread
b. The Piezomed instrument detection cannot detect third-party instruments
-
How long is the useful service life of the instrument detection?
The instrument detection is located in the LED socket and is automatically replaced at the same time as the LED socket.
-
How does it work to wirelessly control the Implantmed and the Piezomed?
W&H introduces a new concept in oral surgery and implantology. Via dongles (receiver), which can be connected to the unit by the doctor, the wireless foot control S-NW communicates with the surgical units of W&H.
-
Piezomed module: Does the module also work with OEM devices?
Yes, but different software updates are necessary.
-
Piezomed module: Can devices be retrofitted on the market?
Yes, all SI-10XX devices.
-
Piezomed module: Are the handpieces of the modules compatible with the SA-320?
No, not compatible.
-
Piezomed module: What are the main differences between the Piezomed Plus module
and Piezomed Classic module?
Features Piezomed Plus module Piezomed Classic module Power 24 watts 18 watts Light LED no light Tip detection yes no Colour black white Cable length 1.8 m and 3.5 m 1.8 m -
Piezomed module: Compatibility of the handpieces with Piezomed Plus module
and Piezomed Classic module.
Handpieces Piezomed Plus module Piezomed Classic module SA-40 yes yes SA-40 L (light version) yes no SA-320 no no -
Piezomed module: Is it possible to upgrade every existing product on the market?
Yes, Implantmed Plus from serial no. SN01001.
-
Can I use tips from competitors?
No, thanks to the automatic tip detection and the special design, W&H offers a closed system.
-
Piezomed module: Can I use the same Piezomed tips for SA-320 and the new modules as well?
Yes.
-
How long is the useful service life of the LED?
-
Elcomed (10)
-
Can handpieces from another manufacturer be used on a W&H surgical motor (implantmed / elcomed)?
For applications which do not require accurate torque (drilling, grinding,…), any handpiece with ISO 3964 connection can be used. Please be informed that hand pieces from other manufacturers might not be able to handle the torque of your unit/motor and therefore might get damaged.
-
Can other brands of tubing be used?
Not recommended. Due to different specifications (material, dimensions) it could cause problems like leakage, insufficient cooling or even damage the unit. For propper tubing see "correlation of tubing" or user manual.
-
How to insert irrigation tube?
Please find a detailed description in the user manual at chapter "Starting operation - General".
-
How can the unit / motor be cleaned / sterilized?
Please find a detailed description in the user manual at chapter "Hygiene and Maintenance".
-
Why doesn't the light on my LED G contra-angle handpiece work when I am positioning an implant?
The generator in the contra-angle handpiece requires at least 300 rpm to produce enough electricity for the LED. The rotary speed is between 15 and 50 rpm when positioning an implant.
-
Can my Implantmed/Elcomed motor be processed via thermal washer disinfector?
Yes, if the thermal washer disinfector has a drying cycle and the motor bears the sign that it is "Thermo washer disinfectable".
-
Can my Implantmed/Elcomed motor be processed in a vacuum sterilizer with the protection cap in place.
No. The new generation of motors have to be sterilized without cap.
-
Does my W&H unit need to be serviced.
Regular servicing of function and safety including the accessories is necessary and should be carried out at least once every three years, unless shorter intervals are prescribed by law. Detailed information can be found in the user manual at chapter "Servicing".
-
What types of tubing can I use on my device?
For propper tubing see "correlation of tubing" or user manual.
-
Does the motor need to be greased?
No, the Implantmed/Elcomed motor does not need greasing. The motor bearings require no maintenance. Any additional greasing would reduce the service life of the bearings.
-
Can handpieces from another manufacturer be used on a W&H surgical motor (implantmed / elcomed)?
Hånd- og vinkelstykker
-
Vinkelstykker (10)
-
Can LED instruments with generator be sterilized?
Yes, W&H handpieces with LED and generator are both sterilizable and thermodisinfectable .
-
Which material is used at W&H surgical instruments?
Internal and external parts are of high quality stainless steel (including the contra-angle handpiece head).
-
How to dismantle the handpiece
See detailed description in user manual or Video.
-
Why doesn't the light on my LED G contra-angle handpiece work when I am positioning an implant?
The generator in the contra-angle handpiece requires at least 300 rpm to produce enough electricity for the LED. The rotary speed is between 15 and 50 rpm when positioning an implant.
-
Is the removable spray clip of my surgical hand piece sterilizable?
Yes. In order to extend the live span of the clip, W&H recommends to remove the clip from the hand piece and sterilize it separately (hand piece and clip in one pouch but separated).
-
What is the difference between WI-75 and WS-75?
Both assure maximum performance. The WS-75 can be dismantled and therefor it can be cleaned more thoroughly.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can our customers disassemble the new contra-angles
Yes. At W&H the letter "S" always communicates, that the surgical contra-angle is disassemble. The letter "I" communicates, that the surgical contra-angle cannot be disassembled (e.g. WS-75 L vs. WI-75)
-
Can the new contra-angles be attached to old motors?
No. In order to deliver the best balance we shortened the inner coupling length. Therefore contra-angles with only an L suffix cannot be attached to old motors
-
Can current contra-angles be attached to new motors?
Yes. All W&H surgical contra-angles and handpieces can be attached to the new motor EM-19.
-
Can LED instruments with generator be sterilized?
-
Håndstykker (9)
-
Can LED instruments with generator be sterilized?
Yes, W&H handpieces with LED and generator are both sterilizable and thermodisinfectable .
-
Which material is used at W&H surgical instruments?
Internal and external parts are of high quality stainless steel (including the contra-angle handpiece head).
-
What is the minimum shaft lenght of the bur for the S-11?
The shaft of the bur should be at least 44mm long (ISO-shaft 104).
-
What is the minimum shaft lenght of the bur for the S-12?
The shaft of the bur should be at least 53mm long.
-
How to dismantle the handpiece
See detailed description in user manual or Video.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can our customers disassemble the new contra-angles
Yes. At W&H the letter "S" always communicates, that the surgical contra-angle is disassemble. The letter "I" communicates, that the surgical contra-angle cannot be disassembled (e.g. WS-75 L vs. WI-75)
-
Can the new contra-angles be attached to old motors?
No. In order to deliver the best balance we shortened the inner coupling length. Therefore contra-angles with only an L suffix cannot be attached to old motors
-
Can current contra-angles be attached to new motors?
Yes. All W&H surgical contra-angles and handpieces can be attached to the new motor EM-19.
-
Can LED instruments with generator be sterilized?
Turbiner
-
Synea (10)
-
Is the new LED turbine from the Synea series e.g. compatible with the existing 924 couplings?
Yes, the new generation of turbines can be operated with both the existing and the new couplings.
-
Which burs can be used with W&H turbines?
Turbines with standard head sizes (models XX-98): 19 mm - 25 mm Turbines with small head sizes (models XX-97): 16 mm - 21 mm
The cutting part diameter for both head sizes is restricted to max. 2 mm. Please observe the technical data provided by the bur manufacturer. When using longer, rotary instruments for special treatment indications, the user must always select the operating conditions correctly in order to avoid placing the user, patient or anybody else at risk. -
How does the new W&H Click & Pull system work? How do I remove a turbine?
Use your thumb and index finger to retract the coupling's retention sleeve. The turbine can now be removed from the coupling without any problems. Pulling forcefully on the dental turbine can cause damage. If excessive force is required to remove the turbine, verify that the retention sleeve on the coupling can be moved without any problems.
-
Why is a 6-hole connection required for the use of an LED turbine?
A 6-hole connection for the turbine tubing is necessary to supply the turbine with electricity via electrical contacts. Other connections do not provide a power supply.
-
What is the maximum length of the rotary instruments?
25 mm for the standard turbine (XX-98 modells) and 21 mm for turbines with small head (XX-97 modells). When using longer rotary instruments the user must ensure by correct selection of the operating conditions, that there is no danger to the user, patient or third parties.
-
I've just bought a new W&H turbine? Should I call a serviceman for my unit to install it?
Yes, it is neccesary to properly adjust the air/water pressure and if applicable the voltage for the turbine.
-
At the moment I have a W&H turbine without light. How can I have a light in my turbine?
If there is electricity at the turbine hose, you need a proper coupling and a turbine with light (example: RQ-24 and TA-97 LED). If no ectricity is available, the simplest way is to buy W&H Alegra LED G turbine which has a generator and LED (light-emitting diode) built in. It doesn't require any other investment in the dental unit.
-
My turbine doesn't have enough power/speed?
Please check drive air pressure on your unit and maintenance of the instruments (Synea with RQ connection = 3 +/- 0,3 bar, with Multiflex® connection 2,5 - 4 bar). For all Alegra turbines the pressure needs to be set between 2,5 - 2,8 bar.
-
Can my W&H turbine be processed via sterilizer?
Yes, providing that your W&H turbine carries the relevant sterilization symbol.
-
Can my W&H turbine be processed via thermal disinfector?
Yes, providing that your W&H turbine carries the relevant thermal disinfection symbol and the thermo disinfector should have a drying cycle.
-
Is the new LED turbine from the Synea series e.g. compatible with the existing 924 couplings?
Tilbehør
-
W&H Osstell ISQ module (7)
-
Can the surgeon upgrade the unit by themself or does the unit need technical service?
The surgeon or dental nurse can easily attach the module to the right side of the Implantmed with the help of the provided quick manual. The module is plug-n-play via a USB connection.
-
Is every new Implantmed compatible with the W&H Osstell ISQ module?
Yes. All control units SI-1023, SI-1015 and SI-1010 can be upgraded with the W&H Osstell ISQ module
-
Is the ISQ scientifically proven?
The Osstell ISQ technology is proven by over 700 scientific studies. These studies have shown, that torque and ISQ correlate.
-
What’s the difference between torque and ISQ?
Torque is a one time, static measurement at the time of placement of the implant and cannot be repeated later on in a non-invasive way. Osstell ISQ measures the lateral stability of the implant and the measurements can be repeated in a non-invasive, dynamic way to monitor the development of osseointegration.
-
Where can I buy the SmartPegs?
Please contact your local W&H partner
The sale of SmartPegs will be defined per country
The surgeon will receive 5 free SmartPegs. A voucher will be provided with the delivery of the product. -
How does the measurement of the ISQ work?
> To measure the ISQ of an implant, you first screw a so called SmartPeg into the placed implant. The SmartPeg, with its magnet on top, works like a small tuning fork. The magnet on the SmartPeg is ‘hit’ with magnetic pulses, from the probe, which makes the SmartPeg vibrate. Due to the stiffness in the interface between the implant surface and the bone the SmartPeg will vibrate accordingly. The more dense the bone is the higher stability and the higher ISQ value.
-
What does ISQ stand for?
The abbreviation ISQ stands for Implant Stability Quotient. It is a value between 1 and 100 and gives the surgeon an insight of the implant stability after placing the implant. With the ISQ value the surgeon can monitor the osseointegration.
-
Can the surgeon upgrade the unit by themself or does the unit need technical service?
-
Piezomed Instrumenter (4)
-
Can Piezomed instruments also be used on third-party devices?
No
a. Different thread
b. Third-party device cannot find the correct resonance point – instrument fails to oscillate -
How long do the instruments last and when should they be replaced?
The useful service life of the instruments is determined by the length of time they are used reprocessing and sterilization cycles have only marginal influence.
a. Instruments need to be replaced
- If there is a loss of power while working
- Where there is visible damage on the working part or the shank of the instrument.
b. Saws must be replaced if
- Teeth are broken off
- Teeth are worn
c. Diamond-coated instruments must be replaced
- As soon as the diamond coating is worn
-
How are the instruments prepared for cleaning?
a. Preparation in the ultrasonic bath:
- Place the instruments in the instrument tray and lower into the ultrasonic bath.
- Use cleaning agents and disinfectants suitable for hand instruments (probes, mirrors, forceps, etc.).
DO NOT USE a bur bath, as this is very aggressive.
- After cleaning in the ultrasonic bath, rinse adequately with water to remove any remnants of the cleaning agent and disinfectant from the coolant channels.
- Blow dry with compressed air after rinsing
- Replace dried instruments in the instrument tray and package for sterilization
b. Mechanical preparation in the cleaning unit and disinfector
- Use the cleaning adapter for cleaning units and disinfectors (see Instructions for Use – Accessories)
- After preparation in the cleaning unit and disinfector, check that the coolant channels are dry and, if necessary, dry again with compressed air. -
Can third-party instruments be used on the Piezomed?
No
a. Different thread
b. The Piezomed instrument detection cannot detect third-party instruments
-
Can Piezomed instruments also be used on third-party devices?
Vedligeholdelsesudstyr
-
Assistina TWIN (7)
-
What does HEPA mean, and what is filtered out of the air?
HEPA is short for 'High Efficiency Particular Air', and the filter is classified as HEPA class E11. This means that the filter retains all particles larger than 1 µm, which includes, among other things, a wide range of bacteria and viruses, suspended particles such as dust and respirable aerosols.
-
How does the user know when the TWIN Care Set needs to be replaced?
The intelligent process monitoring system documents the number of maintenance cycles that have been performed, thereby automatically calculating the number of instruments that can be serviced with the remaining quantity of fluid. The user receives an alert 300 cycles before the cartridge needs replacing (the oil and cleaner LEDs on the top of the unit turn yellow in colour).
-
How often does the TWIN Care Set need to be replaced?
The consumption volumes for each instrument are 0.07 ml of oil and 0.07 ml of cleaning solution. This gives a maximum capacity of 2857 cycles. In reality, the value is around 2800 to 2850 cycles, as initial filling is also required and the fill quantity of the cartridges is subject to certain tolerances that are regulated by the applicable standards.
-
Why is there no rotational lubrication option in the Assistina TWIN?
The oil is atomized in the Assistina TWIN before it is enters the instrument. This ensures full coverage of the gear parts, eliminating the need for technically complex and time-consuming rotational lubrication. However, there is no negative impact on the maintenance quality: the results of a maintenance operation correspond to those achieved with the Assistina 301 Plus.
-
Which processing steps are carried out in the Assistina TWIN?
- Cleaning the spray channels with W&H Activefluid
- Lubricating the rotating parts with W&H Service Oil
- Throughout the maintenance process, potentially contaminated aerosols are extracted from the process chamber and the exhaust air is cleaned using a HEPA filter.
-
Why does the Assistina TWIN have a HEPA filter?
During the maintenance process in the Assistina TWIN, the rotating parts are lubricated with oil and the spray channels rinsed with cleaning solution. These liquids are introduced to the handpieces using compressed air, which produces aerosols that may contain potentially contaminated body fluids (e.g. blood and saliva). In order to ensure that the apparatus poses no risk to the user, the patient or third parties, the law stipulates that aerosols must be prevented from leaking out of the apparatus. We make sure of this by using active suction and the HEPA filter. This HEPA filter must of course be changed regularly, which is why the lifespan of the filter is matched exactly to the capacity of the two cartridges. The HEPA filter therefore reaches the end of its service life once one set of cartridges has been used.
-
What instrument adaptors are available for the Assistina TWIN?
All adaptors that are suitable for the Assistina 3x2/3x3 can also be used in the Assistina TWIN. There is only one exception: the RM/ISO adaptor for the Assistina 3x2/3x3 is not compatible with the Assistina TWIN. Quick Connect adaptors ISO and RM can be used for the Assistina TWIN instead. An overview of all adaptors is available on the product website.
-
What does HEPA mean, and what is filtered out of the air?
-
Assistina 301 plus (4)
-
Correct use of Assistina 301plus.
Please find a detailed description about first startup, regular use and maintenance in the user manual. Furthermore you can find video tutorials about Assistina 301plus.
-
How can I check if my Assistina 301 is working correctly?
Check/replace the O-rings on the motor coupling/adapter (FAQ video Assistina O-rings). Ensure that the Start button is held depressed for at least 2 seconds when starting a programme. When the button is pressed, the green ball should be visible on the right in the functional display (cleaning solution). When the button is then released again, the green ball should be shown on the left. This procedure is also demonstrated in the FAQ video "Using Assistina". A detailed functional check can be found in the "weekly checklist" and in the "Test and functional check" section of the instructions for use.
-
How much air pressure does the Assistina 301 plus require in order to operate properly?
The pressure should be between 4 - 10 bar (58 to 145 psi).
-
How often should the filter in the supply air be changed (Assistina 301 plus)?
Annually or if the filter is full of debris (visual check - filter should be white). Please see video about changing the filter).
-
Correct use of Assistina 301plus.
-
Assistina (3)
-
Correct use of Assistina 301plus.
Please find a detailed description about first startup, regular use and maintenance in the user manual. Furthermore you can find video tutorials about Assistina 301plus.
-
How much air pressure does the Assistina 301 plus require in order to operate properly?
The pressure should be between 4 - 10 bar (58 to 145 psi).
-
How often should the filter in the supply air be changed (Assistina 301 plus)?
Annually or if the filter is full of debris (visual check - filter should be white). Please see video about changing the filter).
-
Correct use of Assistina 301plus.
-
Assistina ONE (6)
-
How do you replace the ONE Care Set?
The two cartridges and the filter can be changed easily without any tools. The fluid is supplied via tapping needles in the bottom of the cartridge spaces that penetrate the cartridge membrane, thereby drawing the fluid out of the cartridge. The two cartridges have different shapes to ensure that the two fluid types cannot be mixed up. In addition, it is only possible to install the HEPA filter incorrectly by using force, which is not necessary.
-
What instrument adaptors are available for the Assistina One?
All adaptors that are suitable for the Assistina 3x2/3x3 can also be used in the Assistina One. There is only one exception: the RM/ISO adaptor for the Assistina 3x2/3x3 is not compatible with the Assistina One. Quick Connect adaptors ISO and RM can be used for the Assistina One instead. An overview of all adaptors is available in the product website.
-
Why does the Assistina One have a HEPA filter?
During the maintenance process in the Assistina One, the rotating parts are lubricated with oil, and the spray channels rinsed with cleaning solution. These liquids are introduced to the handpieces using compressed air, which produces aerosols that may contain potentially contaminated body fluids (e.g. blood and saliva). In order to ensure that the apparatus poses no risk to the user, the patient or third parties, the legislation stipulates that aerosols must be prevented from leaking out of the apparatus. We make sure of this by using active suction and the HEPA filter. This HEPA filter must of course be changed regularly, which is why the lifespan of the filter is matched exactly to the capacity of the two cartridges. The HEPA filter therefore reaches the end of its service life once one set of cartridges has been used.
-
How does the user know when the ONE Care Set needs to be replaced?
The ONE Care Set must be replaced when the cartridges are empty. The levels can be checked thanks to the device's fill level indicators, located on the right and left sides.
-
How often does the ONE Care Set needs to be replaced?
The consumption volumes for each instrument are 0.07 ml of oil and 0.07 ml of cleaning solution. This gives a maximum capacity of 2857 cycles. The value is around 2800 to 2850 cycles, as initial filling is also required and the fill quantity of the cartridges is subject to certain tolerances that are regulated by the applicable standards.
-
Why there is no rotational lubrication option in the Assistina One?
The oil is atomized in the Assistina One before it enters the instrument. This ensures full coverage of the gear parts, eliminating the need for technically complex and time-consuming rotational lubrication. However, there is no negative impact on the maintenance quality.
-
How do you replace the ONE Care Set?
Trådløse enheder
-
Proxeo Twist trådløs (8)
-
How long is the battery life of the Proxeo Twist Cordless?
The rechargeable battery normally lasts for around one working day (8–12 patients with a polishing duration of 4–6 minutes [actual time the foot control is ‘on’]).
-
Are there any recommendations for charging the Proxeo Twist Cordless’ rechargeable battery?
Yes, we recommend charging the battery from time to time (e.g. during the lunch break), in order to achieve optimal performance.
-
How long does the foot control’s rechargeable battery last?
The rechargeable battery lasts approx. two months based on average use (5 working days with 8–12 patients/day for 4–6 minutes).
-
Is it possible to thermally disinfect and sterilize the Proxeo Twist Cordless?
The Proxeo Twist Cordless can be fully disinfected by wiping down. The handpiece sleeve can be sterilized and thermally disinfected. The foot control can also be disinfected by wiping down. You will find further information on this in the Instructions for use 50928.
-
Which prophy angle cups can be used with the Proxeo Twist Cordless?
The Proxeo Twist prophy angle cups from W&H have some outstanding benefits. You can, however, use all conventional prophy angle cups with the Doriot system.
-
How can I connect the Proxeo Twist Cordless with the foot control?
You will find further information on this in the Instructions for use 50928.
-
Should I always call my service department first before I send devices/instruments for repair?
If you are unsure of the cause of the problem, it is better to contact your service department first and talk to a technician. Together with the technician, malfunctions can sometimes be resolved or it may be determined which items need to be sent for repair.
-
Is it the same rechargeable battery as a smartphone and what is the rechargeable battery’s capacity in comparison to the lifespan of the medical device? With a smartphone, I have to charge more often the older it gets.
The Proxeo Twist Cordless has a Li-ion rechargeable battery (like many smartphones). That said, you can’t compare the Proxeo Twist Cordless with a smartphone in this context. Smartphones have apps and updates installed on them which require more processing capacity and are also updated in the background. This shortens the life of the smartphone’s rechargeable battery and means it needs to be recharged more often, even though it only loses a small amount of capacity.
-
How long is the battery life of the Proxeo Twist Cordless?
Hånd- og vinkelstykker
-
Proxeo (1)
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Can my W&H handpiece be processed via sterilizer?
-
Proxeo TWIST (3)
-
Can my W&H handpiece be processed via thermal disinfector and sterilizer.
Yes, providing that your W&H instrument carries the relevant thermal disinfection and sterilization symbol and the thermo disinfector should have a drying cycle.
-
Can my W&H handpiece be processed via sterilizer?
Yes, providing that your W&H handpiece carries the relevant sterilization symbol.
-
Should I call the service station before sending equipment for repair?
If you are not sure what the cause of the problem is, it is better to call the service station and talk to a technician. Together with the technician it can be determined what needs to be sent in and sometimes it is possible to even solve the problem right away.
-
Can my W&H handpiece be processed via thermal disinfector and sterilizer.
Vandbehandling
-
Multidem (2)
-
Which water quality should I use for my sterilizer?
Answer: The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardized. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use).
-
Can the empty Multidem C27 cartridge be returned or must this be disposed of yourself (household waste)?
The products must be disposed in accordance with current local, regional and country regulations; see the disposal considerations (Section 11) contained in the Product Safety Information Sheet.
-
Which water quality should I use for my sterilizer?
-
Osmo (1)
-
Which water quality should I use for my sterilizer?
Answer: The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardized. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use).
-
Which water quality should I use for my sterilizer?
Piezo Scaler
-
Proxeo Ultra (14)
-
Can I treat patients with cardiac pacemakers with W&H Piezo scalers?
Yes, the compatibility of these devices has been assessed in application.
-
How long does the foot control’s rechargeable battery last?
The rechargeable battery lasts approx. two months based on average use (5 working days with 8–12 patients/day for 4–6 minutes).
-
Is the handpiece thermo washer disinfectable?
Yes, as long as your handpiece or contra-angle handpiece carries the mark “thermo washer disinfectable”.
-
Should I always call my service department first before I send devices/instruments for repair?
If you are unsure of the cause of the problem, it is better to contact your service department first and talk to a technician. Together with the technician, malfunctions can sometimes be resolved or it may be determined which items need to be sent for repair.
-
Are W&H tips (e.g. 1U) compatible with the new W&H high-speed mounting system?
No, they are not compatible because they have different threads.
-
How do I know which tip goes with which handpiece?
There are symbols on the tip, the tip changer and the handpiece identifying which system they belong to.
-
Are the new handpieces compatible with the PA-123/PA-115?
No, they are not compatible.
-
Are the new handpieces (PB-5 L, PB-5 L S, PB-5 L Q) compatible with the earlier generation of Piezo scalers?
No, they are not compatible.
-
Can the handpiece be prepared in the sterilizer?
Yes, as long as your handpiece or contra-angle handpiece carries the mark “sterilizable up to the stated temperature”.
-
Can I prepare the handpiece in universal hygiene devices (DAC-Universal)?
Yes.
-
How can I recharge the wireless foot control (PB-530)?
The wireless foot control can be recharged by plugging the power pack directly into the device.
-
How can I connect the wireless foot control with the Proxeo ULTRA PB-530?
Set the power regulator to “OFF”. Connect the control unit and the foot control using the cable. Press and hold the function button for 5 seconds.
-
Can I disinfect the tank by wiping it down?
Yes.
-
Can I disinfect the tank using the thermo washer?
No.
-
Can I treat patients with cardiac pacemakers with W&H Piezo scalers?
Autoklaver
-
Lara (6)
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines.
-
No cycles are stored in the cycle history menu
POSSIBLE CAUSE
An electronic board was replaced when serviced.
SOLUTION
None. The memory of the old board cannot be restored. You are advised to Save the history regularly onto the USB pen drive. -
Warning about programmed maintenance.
POSSIBLE CAUSE
The maintenance program needs to be carried out.
SOLUTION
Contact an authorized W&H service partner. -
Warning about saving to the USB (HTML and SCL files)
POSSIBLE CAUSE
The USB mass storage is disconnected or full.
SOLUTION
Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner -
Warning about consumable replacement.
POSSIBLE CAUSE
A consumable needs to be replaced.
SOLUTION
Order the requested consumable (door seal, dust filter, bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. -
What is the maximum noise level of the Lara sterilizer?
The max. noise level of Lara is 66.9 dB.
-
Do I need to carry out periodic testing on my W&H Sterilizer?
-
Lisa Mini (12)
-
Whom should I get to validate and maintain my sterilizer?
Validation shall be carried out by an official authority. Maintenance tasks shall be carried out by a qualified and authorised technician.
-
Does my sterilizer need any periodic qualification/validation?
Validation generally comprises a range of tests defined in internationally recognised standards. These tests begin for every device with the factory tests at W&H. In addition, corresponding documentation is included with your device. Prior to commissioning the device at your premises, the technician performs further checks to ensure the device functions correctly. Then, depending on the national/regional laws and directives, the device must be subjected to a so-called renewed performance qualification at regular intervals. For further information, please contact your dealer.
-
How long should I archive the documentation of regular tests?
Keep the results according to your country/regional specific laws and guidelines. Those records can be of assistance in case of service as well.
-
Do I need to document the results of all periodic tests?
W&H recommends recording all periodic tests, repairs and modifications on the sterilizer in a logbook including date and signature of the person who carried out the test or other activity. Compulsories of recording is regulated by local/national guidelines.
-
What is validation?
Validation is a testing procedure carried out by an official authority, proving the effectiveness of your sterilizer. Compulsories and frequency of validation are regulated by local/national guidelines.
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines.
-
Which water quality should I use for my sterilizer?
Answer: The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardized. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use).
-
Should I pouch/wrap my instruments and how long can I store them after sterilization?
To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer.
-
What is the maximum noise level of the Lisa Mini sterilizer?
The max. noise level of Lisa Mini is 64 dB.
-
Is Lisa Mini requiring compressed air connection?
Lisa Mini requires compressed air connection to perform sterilization cycle. The compressed air specifications are available in the Instruction for Use, chapter “Technical data”.
-
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu.
-
How do I clean the water tank and what products should I use?
Water tank cleaning instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu.
-
Whom should I get to validate and maintain my sterilizer?
-
Lina (15)
-
How do I clean the water tanks and what products should I use?
Information and videos relating to the cleaning of the water tanks and cleaning products can be found on the Video Tutorial section of this web site.
-
Whom should I get to validate and maintain my sterilizer?
Validation shall be carried out by an official authority. Maintenance tasks shall be carried out by a qualified and authorised technician.
-
Does my sterilizer need any periodic qualification/validation?
Validation generally comprises a range of tests defined in internationally recognised standards. These tests begin for every device with the factory tests at W&H. In addition, corresponding documentation is included with your device. Prior to commissioning the device at your premises, the technician performs further checks to ensure the device functions correctly. Then, depending on the national/regional laws and directives, the device must be subjected to a so-called renewed performance qualification at regular intervals. For further information, please contact your dealer.
-
How long should I archive the documentation of regular tests?
Keep the results according to your country/regional specific laws and guidelines. Those records can be of assistance in case of service as well.
-
Do I need to document the results of all periodic tests?
W&H recommends recording all periodic tests, repairs and modifications on the sterilizer in a logbook including date and signature of the person who carried out the test or other activity. Compulsories of recording is regulated by local/national guidelines.
-
What is validation?
Validation is a testing procedure carried out by an official authority, proving the effectiveness of your sterilizer. Compulsories and frequency of validation are regulated by local/national guidelines.
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines.
-
Which water quality should I use for my sterilizer?
Answer: The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardized. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use).
-
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu (only for Lisa); • in the Consumables & Accessories and Video Tutorial section of this web site
-
No cycles are stored in the cycle history menu
POSSIBLE CAUSE
An electronic board was replaced when serviced.
SOLUTION
None. The memory of the old board cannot be restored. You are advised to Save the history regularly onto the USB pen drive. -
What is the maximum noise level of the sterilizer?
The silent running noise at a maximum level of 66.5 dB(A) guarantees a quiet working environment.
-
Should I pouch/wrap my instruments and how long can I store them after sterilization?
To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer.
-
Warning about programmed maintenance.
POSSIBLE CAUSE
The maintenance program needs to be carried out.
SOLUTION
Contact an authorized W&H service partner. -
Warning about saving to the USB (HTML and SCL files)
POSSIBLE CAUSE
The USB mass storage is disconnected or full.
SOLUTION
Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner -
Warning about consumable replacement.
POSSIBLE CAUSE
A consumable needs to be replaced.
SOLUTION
Order the requested consumable (door seal, dust filter, bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use.
-
How do I clean the water tanks and what products should I use?
-
Lexa (1)
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines.
-
Do I need to carry out periodic testing on my W&H Sterilizer?
-
Lara XL (7)
-
How do I clean the water tanks and what products should I use?
Information and videos relating to the cleaning of the water tanks and cleaning products can be found on the Video Tutorial section of this web site.
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines.
-
Which water quality should I use for my sterilizer?
Answer: The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardized. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use).
-
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu (only for Lisa); • in the Consumables & Accessories and Video Tutorial section of this web site
-
Warning about saving to the USB (HTML and SCL files)
POSSIBLE CAUSE
The USB mass storage is disconnected or full.
SOLUTION
Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner -
Warning about consumable replacement.
POSSIBLE CAUSE
A consumable needs to be replaced.
SOLUTION
Order the requested consumable (door seal, dust filter, bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. -
What is the maximum noise level of the Lara sterilizer?
The max. noise level of Lara is 66.9 dB.
-
How do I clean the water tanks and what products should I use?
-
Lisa (25)
-
How do I clean the water tanks and what products should I use?
Information and videos relating to the cleaning of the water tanks and cleaning products can be found on the Video Tutorial section of this web site.
-
Lisa 17/22
At the end of the cycle, there is residual water in the chamberPOSSIBLE CAUSE
Sterilizer not properly levelled
SOLUTION
Properly level the surface that the sterilizer is placed on.
POSSIBLE CAUSE
Overloaded chamber
SOLUTION
Comply with the maximum load weight limits for each type of load. Always use the chamber rack for trays and cassettes.
POSSIBLE CAUSE
Chamber filter clogged
SOLUTION
Remove and clean the chamber filter.
POSSIBLE CAUSE
Chamber filter cap mispositioned
SOLUTION
Mount the chamber filter cap properly (see chapter MAINTENANCE)
POSSIBLE CAUSE
Load incorrectly placed
SOLUTION
Follow the recommendations as listed in ANNEX 2. -
Lisa 17/22
Corrosion or spots on instrumentsPOSSIBLE CAUSE
Tap water on instruments when placed in the sterilizer
SOLUTION
Ensure that instruments are dry before they are placed in the sterilizer.
POSSIBLE CAUSE
Use of water of poor quality or water containing chemical substances
SOLUTION
Drain both water tanks. Use good quality water (see APPENDIX 7).
POSSIBLE CAUSE
Organic or chemical residues on the instruments
SOLUTION
Clean, rinse and dry instruments before placing them in the sterilizer (see APPENDIX 2).
POSSIBLE CAUSE
Chamber, trays, tray rack dirty
SOLUTION
Clean the chamber and wash the chamber furniture
POSSIBLE CAUSE
Contact between instruments of different materials
SOLUTION
Ensure that instruments of different materials do not touch (aluminum, carbon or stainless steel, etc.); place them on different trays or cassettes or pouch them (refer to APPENDIX 2).
POSSIBLE CAUSE
Scale deposits on the chamber
SOLUTION
Clean the chamber and use good quality water (refer to APPENDIX 7). -
Lisa 17/22
When the sterilizer is connected to an automated water supply system: there is no clean water in the tank, but the automatic water filling does not fill the water.POSSIBLE CAUSE
Water fill system not connected
SOLUTION
Connect the water fill system to the sterilizer (see ANNEX 7 for water quality requirements).
POSSIBLE CAUSE
When the water fill system attempted to fill the tank, water was temporarily unavailable
SOLUTION
Since water tank filling is attempted only once in-between cycle execution, this event inhibits water feeding. Switch the sterilizer OFF and then ON again.
Check the external water supply system.
Check for water leaks from the sterilizer.
POSSIBLE CAUSE
Faulty MIN water level sensor in the clean water tank
SOLUTION
Call service. -
Whom should I get to validate and maintain my sterilizer?
Validation shall be carried out by an official authority. Maintenance tasks shall be carried out by a qualified and authorised technician.
-
Does my sterilizer need any periodic qualification/validation?
Validation generally comprises a range of tests defined in internationally recognised standards. These tests begin for every device with the factory tests at W&H. In addition, corresponding documentation is included with your device. Prior to commissioning the device at your premises, the technician performs further checks to ensure the device functions correctly. Then, depending on the national/regional laws and directives, the device must be subjected to a so-called renewed performance qualification at regular intervals. For further information, please contact your dealer.
-
How long should I archive the documentation of regular tests?
Keep the results according to your country/regional specific laws and guidelines. Those records can be of assistance in case of service as well.
-
Do I need to document the results of all periodic tests?
W&H recommends recording all periodic tests, repairs and modifications on the sterilizer in a logbook including date and signature of the person who carried out the test or other activity. Compulsories of recording is regulated by local/national guidelines.
-
What is validation?
Validation is a testing procedure carried out by an official authority, proving the effectiveness of your sterilizer. Compulsories and frequency of validation are regulated by local/national guidelines.
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines.
-
Which water quality should I use for my sterilizer?
Answer: The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardized. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Water treatment devices Multidem and Osmo are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use).
-
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu (only for Lisa); • in the Consumables & Accessories and Video Tutorial section of this web site
-
The sterilizer enters into “Standby mode” immediately after being switched on
POSSIBLE CAUSE
The unit was turned off while in standby mode.
SOLUTION
Press the standby mode button to exit. -
When starting a cycle, the chamber door locks but re-opens immediately. The “Open the door” message appears.
POSSIBLE CAUSE
Door seal not properly placed.
SOLUTION
Ensure that the door seal is correctly inserted around the whole circumference.
POSSIBLE CAUSE
Door jammed by external objects or by the load itself.
SOLUTION
Remove any objects interfering with the chamber door. Check for interference between the door and the chamber furniture and/or the load. -
No cycles are stored in the cycle history menu
POSSIBLE CAUSE
An electronic board was replaced when serviced.
SOLUTION
None. The memory of the old board cannot be restored. You are advised to Save the history regularly onto the USB pen drive. -
The cycle report printer does not work.
POSSIBLE CAUSE
Printer not properly connected or not powered.
SOLUTION
Check the data and power connection to the printer. -
Water is leaking at the front of the sterilizer
POSSIBLE CAUSE
Leaks through the chamber door seal.
SOLUTION
Clean or replace the door seal. Clean the door surface facing the chamber.
POSSIBLE CAUSE
Internal leak.
SOLUTION
Contact an authorized W&H service partner. -
The sterilizer remains switched OFF.
POSSIBLE CAUSE
The main switch or network circuit breaker is OFF.
SOLUTION
Activate the main switch or network circuit breaker (ON).
POSSIBLE CAUSE
No voltage at the socket.
SOLUTION
Check the building electrical supply.
POSSIBLE CAUSE
The power cord is not connected properly.
SOLUTION
Check and connect the power cord properly -
What is the maximum noise level of the sterilizer?
The smooth running noise at a pleasant maximum level of 64.3 dB(A) ensures a high comfort working environment.
-
Should I pouch/wrap my instruments and how long can I store them after sterilization?
To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer.
-
Warning about programmed maintenance.
POSSIBLE CAUSE
The maintenance program needs to be carried out.
SOLUTION
Contact an authorized W&H service partner. -
Warning about saving to the USB (HTML and SCL files)
POSSIBLE CAUSE
The USB mass storage is disconnected or full.
SOLUTION
Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner -
The sterilization (PROCESS) phase of a sterilization cycle was longer than expected.
POSSIBLE CAUSE
The chamber temperature dropped below the minimum threshold and the software performed a successful recovery.
SOLUTION
Wait for cycle completion. If the problem occurs frequently, contact an authorized W&H service partner -
Warning about consumable replacement.
POSSIBLE CAUSE
A consumable needs to be replaced.
SOLUTION
Order the requested consumable (door seal, dust filter, bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. -
What is the maximum noise level of the Lisa sterilizer?
The max. noise level of Lisa 2019 is 64.3 dB.
-
How do I clean the water tanks and what products should I use?
Luft Motor
-
Does the air motor need to be greased?
Yes, the air motor must be greased as described in the "Hygiene and care" section of the instructions for use.